4 .4 Basic Scanning Mode
155
Chapter 4
z
Navigation type 3
Outside of range display example
Inside of range display example
Navigation type 3 displays the pressurization stability .
If the pressurization strength on the right and left sides of the transducer differs the elastic image coloration will
be uneven .
Navigation type 3 displays the pressurization strength on the right and left sides of the transducer . Change the
transducer angle and repeat the pressurization/relaxation to achieve a horizontal line and display the evenly col-
ored elastic image .
Stable Interval
Stable interval display is a function which displays the interval by assuming continuous pressurization has been per-
formed to achieve stable pressure .
When On is selected for the stable interval setting, part of the Cine bar turns green to indicate the stable interval at
the moment the display is frozen .
Searching the stable interval within Cine makes it easier to find the best image for the diagnosis.
IMPORTANT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
• A stable interval assumes stable pressurization has been performed . Not all intervals are appropriate for the diagno-
sis .
• During pressurization or inappropriate pressurization (excessive distortion or lateral movement of the cross-section)
images that are unsuitable for diagnosis may be selected .
• When pressurization is intermittent (i .e . heartbeat, etc .) stable intervals may be undetectable .
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Содержание SONIMAGE HS2
Страница 1: ...EN 01...
Страница 2: ......
Страница 6: ...6...
Страница 7: ...7 Introduction...
Страница 41: ...41 For all product inquiries contact your service representative Support...
Страница 42: ...42...
Страница 43: ...43 Chapter 1 Product Overview...
Страница 68: ...68...
Страница 69: ...69 Chapter 2 Before Examination...
Страница 84: ...84...
Страница 85: ...85 Chapter 3 Starting Examination...
Страница 122: ...122...
Страница 123: ...123 Chapter 4 Diagnosis Mode...
Страница 181: ...181 Chapter 5 Common Function in Ultrasound Images...
Страница 219: ...219 Chapter 6 Reviewing Printing and Transferring Images...
Страница 232: ...232...
Страница 233: ...233 Chapter 7 Maintenance Inspection...
Страница 243: ...243 Chapter 8 Transducer and Biopsy...
Страница 270: ...270...
Страница 271: ...271 Chapter 9 MI TI...
Страница 283: ...283 Chapter 10 Acoustic Output...
Страница 286: ...286...
Страница 287: ...287 Chapter 11 Others...
Страница 305: ...305...
Страница 306: ...306...
Страница 307: ......