AD3435_en
|
2017-11-17
5–1
5
Specifications
This section identifies the system specifications and the site requirements to operate the
imager.
Equipment Specifications
Operating Requirements
•
Allow 31 cm (12 in.) clearance around the top, sides, and back of the imager. This
space is required to let the imager perform normal operator functions.
•
The table or counter must be level (within 1°) and capable of supporting the imager’s
weight.
The surface
of the table or counter
must extend beyond the imager
dimensions by at least 31 cm
(12 in.). In addition, there must be space in front of the
imager to allow access in
the event of a service call.
•
The recommended table or counter height is 71–76 cm (28–30 in.).
•
Place the imager in an area with good ventilation. A small, confined room is not
recommended.
•
Avoid placing the imager in direct or excessive sunlight (for example, near a large
window)
or in an environment with dust, dirt, or airborne chemicals.
Unpacked
Height
66.6 cm (26.2 in.)
Width
62.6 cm (24.6 in.)
Depth
64.9 cm (25.5 in.)
Weight
79 kg (175 lb)
Note
The approximate weight of a 35 x 43 cm (14 x 17 in.) film
cartridge is 3.5 kg (7.7 lb).