E-88
Assessments
4
Turn
B
to select “TOLERANCE”, then press
A
.
The currently set tolerances will be
displayed.
5
If they are satisfactory, turn
B
to select “BREAK”, then press
A
.
The measurement screen will ap-
pear.
6
Place the specimen measuring port to the specimen, and then press
C
(MEAS.
button).
The color difference, direction and amount of tint and brightness deviation from the target color will be dis-
played. In addition, on the right of “ E*” value, a pass/fail judgment result made based on the specified
elliptical tolerances will be displayed.
• The color space to be used as the assessment axis differs depending on the selected color space.
When “L*C*h, E”, “L*C*h, CMC1:1”, “L*C*h, CMC2:1” or “L*C*h, CIE94” is selected:
Assessment will be made based on the direction of tint and lightness deviation and color difference
from the target color in the L*C*h axis. When a color space other than those given above is selected,
assessment will be made based on the direction of tint and lightness deviation and color difference
from the target color in the L*a*b* axis.
M
emo
• Assessment results differ as shown in the table in page E-84 depending on the type of specified color difference
tolerances and pass/fail judgment result.
Содержание CM-2600d
Страница 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Страница 11: ...E 9 Chapter 1 Before Using the Instrument...
Страница 27: ...E 25 Chapter 2 Preparation for Measurement...
Страница 62: ...E 60 Setting Color Difference Tolerances...
Страница 63: ...E 61 Chapter 3 Measurement...
Страница 79: ...E 77 Chapter 4 Other Functions...
Страница 108: ...E 106 TASK Mode...
Страница 109: ...E 107 Chapter 5 Troubleshooting...
Страница 114: ...E 112 Troubleshooting...
Страница 115: ...E 113 Chapter 6 Appendix...
Страница 122: ...E 120 Dimensions Dimensions 193 96 96 69 mm The measurement area selector is not available with the CM 2500d...
Страница 123: ...E 121 Dimensions...
Страница 126: ......
Страница 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...