E-17
Preparation
Cleaning Each Part
This section explains how to clean the White Calibration Plate, Target Mask and inside of the integrating
sphere.
White Calibration Plate
Gently wipe off dirt with a soft dry cloth. If dirt is difficult to remove, dampen a cloth with commercially
available lens cleaning liquid and wipe. Then remove the liquid with a cloth dampened with water, and
leave it to dry.
N
ote
Take care not to scratch the White Calibration Plate.
Target Mask
Use a blower to remove dirt and dust from the Target Masks.
N
ote
Do not touch the inner surface of the Target Masks with your fingers or wipe them with a cloth. If the Target
Masks are so dirty that dirt cannot be removed with a blower, contact the nearest KONICA MINOLTA SENSING-
authorized service facility.
Inside the Integrating Sphere
M
emo
For attaching/removing a Target Mask, refer to “Attaching/Removing a Target Mask” (page E-14).
1. To prevent dust and dirt entering the optics sec-
tion from the integrating sphere, slide the view-
finder lever to open the viewfinder.
2. Remove the Target Mask.
3. Use a blower to remove dirt and dust from the in-
tegrating sphere.
N
ote
Do not touch the white-coated inner surface of the integrating
sphere, wipe it with a cloth or put any object inside it. If the surface
is so dirty that dirt cannot be removed with a blower, contact the
nearest KONICA MINOLTA SENSING-authorized service facility.
Содержание CM-2600d
Страница 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Страница 11: ...E 9 Chapter 1 Before Using the Instrument...
Страница 27: ...E 25 Chapter 2 Preparation for Measurement...
Страница 62: ...E 60 Setting Color Difference Tolerances...
Страница 63: ...E 61 Chapter 3 Measurement...
Страница 79: ...E 77 Chapter 4 Other Functions...
Страница 108: ...E 106 TASK Mode...
Страница 109: ...E 107 Chapter 5 Troubleshooting...
Страница 114: ...E 112 Troubleshooting...
Страница 115: ...E 113 Chapter 6 Appendix...
Страница 122: ...E 120 Dimensions Dimensions 193 96 96 69 mm The measurement area selector is not available with the CM 2500d...
Страница 123: ...E 121 Dimensions...
Страница 126: ......
Страница 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...