E-46
3
Turn
B
to select “ZERO”.
4
Direct the specimen measuring port into the air.
1m or more
• Never place the specimen measuring port toward a light source.
•
Keep the specimen measuring port more than 1m away from any reflective items (hands, desks,
walls etc.)
.
5
Make sure that
is displayed, then press
C
(MEAS. button).
Zero calibration will be performed.
When zero calibration is complete,
the <CALIBRAITION> screen will
reappear.
M
emo
• Use of the optional Zero Calibration Box ensures proper zero calibration.
If Zero Calibration Box is used, remove
“Measuring Base”.
• If
does not appear due to the auto power save function, this may be due the fact that it can take time
to start zero calibration after
C
is pressed.
N
ote
• When zero calibration is complete, perform white calibration. Measurement cannot be performed if
white calibration has not been completed.
• Even though white calibration has been performed after the power is turned ON, if zero calibration has
also been performed after that, white calibration must be performed again.
Содержание CM-2600d
Страница 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Страница 11: ...E 9 Chapter 1 Before Using the Instrument...
Страница 27: ...E 25 Chapter 2 Preparation for Measurement...
Страница 62: ...E 60 Setting Color Difference Tolerances...
Страница 63: ...E 61 Chapter 3 Measurement...
Страница 79: ...E 77 Chapter 4 Other Functions...
Страница 108: ...E 106 TASK Mode...
Страница 109: ...E 107 Chapter 5 Troubleshooting...
Страница 114: ...E 112 Troubleshooting...
Страница 115: ...E 113 Chapter 6 Appendix...
Страница 122: ...E 120 Dimensions Dimensions 193 96 96 69 mm The measurement area selector is not available with the CM 2500d...
Страница 123: ...E 121 Dimensions...
Страница 126: ......
Страница 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...