Operation of DR Detector
33
3�2�3
DR Detector orientation
Change the orientation of the DR Detector according
to the exposure body part.
Expose with the triangular mark upward when expos-
ing in portrait, and with the triangular mark to the side
when exposing in landscape.
IMPORTANT
•
The side with the triangular mark is the exposure surface.
When exposing in portrait
X-ray
X-ray device
The triangular mark is upward
When exposing in landscape
X-ray
X-ray device
The triangular
mark is to the
left or right
HINT
•
The blue lines also helps you identify the top-bottom orien-
tation of the DR Detector.
•
To carry out exposures in portrait orientation, turn the DR
Detector so that its blue lines comes to the bottom position.
Blue lines
3�2�4
Precautions for exposure
Exposure area
IMPORTANT
•
If the irradiation field is larger than DR Detector when you
take an image, an artifact may occur due to the X-ray back-
scatter effect. Adjust the irradiation filed to fit within the im-
aging range.
•
The size lines of DR Detector indicate the exposure
size.
•
During exposure, place the exposure body part with-
in the white lines on the four corners.
White lines
Blue lines
Size lines
Содержание AeroDR SKR 4000
Страница 1: ...EN 02...
Страница 2: ......
Страница 8: ...8...
Страница 9: ...9 Chapter 1 Safety Precautions Warnings...
Страница 22: ...22...
Страница 23: ...23 Chapter 2 Product Overview...
Страница 29: ...29 Chapter 3 General Operations...
Страница 42: ...42...
Страница 43: ...43 Chapter 4 Troubleshooting...
Страница 46: ...46...
Страница 47: ...47 Chapter 5 Maintenance...
Страница 50: ...50...
Страница 51: ...51 Chapter 6 Specifications...
Страница 56: ...56...
Страница 57: ...57...
Страница 58: ...58...
Страница 59: ......