background image

36

3�3 

Inserting and removing the Battery

IMPORTANT

 •

Handle the Battery carefully when inserting and removing 

the Battery. The battery connector port may be damaged.

 •

Turn off the device when inserting and removing the Bat-

tery

3�3�1 

Insertion

Ensure that the Status LED of the DR De-

tector is turned off.

Status LED 

 

(green/orange)

Align the insertion direction indication 

mark on the Battery with the DR Detector 

battery connector port, and then insert the 

Battery�

Insertion direction indication mark

Battery connector port

Slide the battery lock and lock the Battery�

Battery lock

Battery lock

HINT

 •

When a Battery charged more than 11 % is inserted, the 

DR Detector starts up and the Battery LED (green) turns 

on

3�3�2 

Removal

Ensure that the Status LED of the DR De-

tector is turned off.

Status LED 

 

(green/orange)

Slide the battery lock to unlock the Bat

-

tery�

Battery lock

Battery lock

Remove the Battery.

Содержание AeroDR SKR 4000

Страница 1: ...EN 02...

Страница 2: ......

Страница 3: ...2 2 Component names and functions 26 2 2 1 DR Detector 26 2 2 2 Battery 27 2 2 3 Battery Charger 27 2 2 4 DC Power Cable 28 Chapter 3 Chapter 3 General Operations 29 3 1 Startup and shutdown 30 3 1 1...

Страница 4: ...r 6 Specifications 51 6 1 Specifications 52 6 1 1 DR Detector 52 6 1 2 Battery 53 6 1 3 Battery Charger 53 6 1 4 AC Adapter 54 6 1 5 DC Power Cable 54 6 1 6 SKR 4000 54 6 2 Characteristics and perform...

Страница 5: ...ative This operation manual provides instructions on the basic functions for operation of this device Those operating this device for the first time should read this manual beforehand Also store this...

Страница 6: ...to the console the image processing controller with wireless communication There are two ways to supply the power to the DR Detector from removable Battery and wired connection through the AC Adapter...

Страница 7: ...sume any liability relating to the following events arising due to systems in fected by computer viruses worms and other malicious software potentially harmful to computers hereunderre ferred to as ma...

Страница 8: ...8...

Страница 9: ...9 Chapter 1 Safety Precautions Warnings...

Страница 10: ...e Graphic symbols Explanation Indicates that full caution is required when handling this device This indicates to reference to the instruction manual for general information Indicates that it is neces...

Страница 11: ...are there to make sure that the user recognizes potential hazards when operating this device IMPORTANT If a warning label is too dirty or damaged to read contact Konica Minolta technical representati...

Страница 12: ...se follow the rules and regulations of your relevant authorities in the disposal of this product accessories options consumables media and their packing materi als A 15 minute warm up operation is req...

Страница 13: ...attery to leak In respect to Battery pay attention to the followings to pre vent heating explosion and fire Do not expose Battery to strong impact dropping the battery from an elevated location or thr...

Страница 14: ...e between each DR Detector for at least 30 cm When using the DC Power Cable observe the following Remove the cable by holding the connector housing Do not let the cable get pinched by doors and do not...

Страница 15: ...the limits for medical devices in IEC 60601 1 2 2014 These limits are designed to provide reasonable pro tection against harmful interference in a typical medi cal installation The device generates u...

Страница 16: ...transmitters frequen cy Wireless standard IEEE802 11a b g n ac Frequency 2 4G 2 412 2 4835 GHz 5G 5 15 5 85 GHz Modulation 802 11b CCK DQPSK DBPSK 802 11a g 64 QAM 16 QAM QPSK BPSK 802 11n 64 QAM 16...

Страница 17: ...ess communications equipment Electrical fast transients bursts IEC 61000 4 4 2 kV input d c power port 2 kV input d c power port 1 kV signal input output parts port 100 kHz repetition frequency Surges...

Страница 18: ...ency MHz Modulation Immunity test level V m 385 Pulse modulation 18 Hz 27 450 Frequency modulation 5 kHz deviation 1 kHz sine 28 710 Pulse modulation 217 Hz 9 745 780 810 Pulse modulation 18 Hz 28 870...

Страница 19: ...y additional research DR Detector has been tested and found to comply with FCC IC radiation expo sure limits set forth for controlled equipment and meets the FCC radio frequency RF Exposure Guidelines...

Страница 20: ...g to compulsory certification of Electrical Equipment Subject to Sanitary Surveillance by ANVISA RDC No 27 dated 21 06 2011 and Nor mative Instruction No 11 dated 16 12 2014 The highest SAR value repo...

Страница 21: ...DR Detector If you directly apply or spray disinfecting solution on the DR Detector the solution will enter the instrument through exterior gaps causing a failure Dampen a lint free soft cloth with di...

Страница 22: ...22...

Страница 23: ...23 Chapter 2 Product Overview...

Страница 24: ...ss point Communicates with the DR Detector Provided by a general purpose product 5 Image processing controller Controls the reception management and output of image data Up to 2 DR Detectors can be re...

Страница 25: ...w of this device Connection example 2 X ray Wireless connection 1 DR Detector Wireless connection 1 DR Detector Exposure room 2 Battery Charger 3 DC Power Cable X ray device operation panel 5 Image pr...

Страница 26: ...eless modes 3 Triangular mark Indicates the direction to place the DR Detector in during exposure 4 Antenna Displays the place where a wireless antenna is attached 5 Exterior Protects the internal par...

Страница 27: ...1 Battery connector port Connects to the DR Detector or Battery Charger 2 Insertion direction indication mark Indicates the insertion direction of the Battery 2 2 3 Battery Charger 1 2 Number Name Fu...

Страница 28: ...tions 2 2 4 DC Power Cable 1 2 3 4 Number Name Functions 1 Wired connection connector Connects to the DR Detector 2 LED Display the DC Power Cable status 3 AC Adapter Used to supply power to the DC Po...

Страница 29: ...29 Chapter 3 General Operations...

Страница 30: ...ting a charged Battery to the DR Detector Connecting the DC Power Cable to the DR Detector and the AC Power Cable to the power outlet Battery Charger 1 Connect the AC Power Cable to a wall out let and...

Страница 31: ...31 Battery Charger 1 Disconnect the AC Power Cable from the wall outlet and confirm that the Power LED green is off Power LED green DC Power Cable 1 Disconnect the AC Power Cable from the wall outlet...

Страница 32: ...is dis played on the image processing control ler HINT When inserting the DR Detector to the wall stand or table an exposure operation can be performed with the DC Pow er Cable connected Press the Pow...

Страница 33: ...riangular mark is to the left or right HINT The blue lines also helps you identify the top bottom orien tation of the DR Detector To carry out exposures in portrait orientation turn the DR Detector so...

Страница 34: ...y Grid ratio Conver gence distance Angle error AeroDR NS 1417 40 lp cm Variety Variety 5 0 or less High dose exposure When high dose exposures are performed continu ously afterimages of the last expos...

Страница 35: ...a location near a de vice that emits electromagnetic wave noise may ap pear on the image When performing exposure turn off that device or move that device away from DR Detector Waterproof performance...

Страница 36: ...rection indication mark on the Battery with the DR Detector battery connector port and then insert the Battery Insertion direction indication mark Battery connector port 3 Slide the battery lock and l...

Страница 37: ...DC Power Cable 6 hours or less if not expose 3 4 1 Charging with Battery Charger IMPORTANT Handle the Battery carefully when inserting and removing the Battery The battery connector port may be damage...

Страница 38: ...Power Cable When charging starts the Battery LED green orange on the DR Detector blinks When charged more than 11 the Battery LED green blinks Battery LED green orange HINT The DR Detector will be cha...

Страница 39: ...ecrease When the Battery is removed if the DC Power Cable is connected and the DR Detector is turned on the Battery LED green turns up LINK LED Display pattern Lighting color Status Shutdown condition...

Страница 40: ...arger Charging LED Power LED Fully charged LED Power LED green Display pattern Status Shutdown condition Operating Charging LED orange and Fully charged LED green Charging LED Fully charged LED Status...

Страница 41: ...ffix the labels only in the recommended places Failure to do so may cause the labels to come off or image uneven ness to occur Position to affix DR Detector identification label XXXXXXXXXXXXXXXX 1 2 H...

Страница 42: ...42...

Страница 43: ...43 Chapter 4 Troubleshooting...

Страница 44: ...e Status LED green does not light Ready is not displayed on the image processing controller System error is occurring If the status LED green does not turn on after approx 10 minutes restart the DR De...

Страница 45: ...ms Did you perform exposure in Ready status Are the exposure conditions correct Was X ray output lowered Was X ray voltage set to a lower value 2 Perform the following procedures Increase the voltage...

Страница 46: ...46...

Страница 47: ...47 Chapter 5 Maintenance...

Страница 48: ...aning the Battery Charger Weekly Full charge of the Battery 6 months 5 1 2 Cleaning IMPORTANT Be careful not to apply any cleaning chemical or liquid onto the wired connection connectors the Power SW...

Страница 49: ...ed it may affect the quality performance and safety of the DR Detector Because the antiseptic solution is a chemical agent adhere to the precautions provided by the pharmaceutical manu facturer If bod...

Страница 50: ...50...

Страница 51: ...51 Chapter 6 Specifications...

Страница 52: ...luding the Battery Pixel size 150 um Image area size 345 6 420 mm 2 304 2 800 pixels AD conversion 16 bit Maximum patient weight Point load 100 kg 40 mm Face load 150 kg effective image area overall D...

Страница 53: ...atus 2 5 hours Without sleep 7 6 hours With sleep 6 1 2 Battery Item Description Product name BATTERY KV made by iRay Technology Co Ltd Battery type Lithium Ion Battery Rated capacity 4180 mAh standar...

Страница 54: ...35 RH to 85 RH ensure no water con densation 35 RH 85 RH 700 hPa to 1060 hPa 700 hPa 1060 hPa Temperature change should be less than 1 C min Pressure change should be less than 1 0197 10 4 Pa min When...

Страница 55: ...87 870 8700 0 0 5 1 1 5 2 2 5 3 0 1000 2000 3000 4000 0 0001 0 001 0 01 0 1 1 10 100 1000 Air Kerma Gy Density Pixel Value Exposure Dose mR P 41 REGIUS210 QR 250 SRO 500 SR G right axis MTF 0 0 0 2 0...

Страница 56: ...56...

Страница 57: ...57...

Страница 58: ...58...

Страница 59: ......

Страница 60: ...ia Private Ltd Office No 201 2nd Floor Atrium 2 Next to Courtyard Marriott Hotel Andheri Kurla Road Chakala Andheri East Mumbai 400093 India TEL 91 22 61916900 Konica Minolta Business Solutions Canada...

Отзывы: