378511/A
63
Optical isolated input signals
330
Ω
I
F
U =1.2V
F
Trig In+
Trig In-
R
E
CD020104_102_001
A
Input from external system
B
Processing Unit input
circuitry
The input current must
be approximately 10 mA.
Depending on your input signal
additional resistance must be
applied to achieve the required
input current. Two examples
are shown to clarify.
Example:
I =
F
»
10mA
330
W
4.5V - 1.2V(U )
F
Using +4.5V input signal the input current will be as required (~10mA). No additional
resistance required.
Example:
R =1080-330=750
E
W
R
=
TOT
=
=1080
W
10mA
12V - 1.2V(U )
F
0.010
10.8
An added resistor of 750Ω and minimum 0.1 Watt must be used.
Note
The input signals must not be negative, i.e. no RS-232 signals can be used for these
inputs.
Optical isolated output signals
The collector current must be approximately 10 mA. A resistor must be used to tune the
collector current depending on your voltage.
Содержание EM 2040C
Страница 1: ...Maintenance manual EM 2040C Multibeam echo sounder...
Страница 2: ......
Страница 3: ...378511 A August 2016 Kongsberg Maritime AS EM 2040C Multibeam echo sounder Maintenance manual...
Страница 8: ...6 378511 A EM 2040C...
Страница 10: ...8 378511 A EM 2040C Topics System description page 9 System diagram page 11 System units page 12 EM 2040C...
Страница 16: ...14 378511 A Troubleshooting Topics BIST Built In Self Test dialog box page 15 EM 2040C...
Страница 71: ...378511 A 69 Hydrographic Work Station outline dimensions Drawing 378828 2 pages Drawing file...
Страница 72: ...70 378511 A Page 2 EM 2040C...
Страница 123: ...378511 A 121 Hydrographic Work Station 50 processing unit 45 48 Remote Control Unit K Rem 73 topside 50 Index...
Страница 124: ...2016 Kongsberg Maritime...