K71000 Automatic PMCC Flash Point Analyzer 

Operation and Instruction Manual 

 
 

K71000-Manual

    

 

-

7-

 

 

Once  analysis  is  complete  the  results  are 
displayed  to  the  screen  and  saved  to  the 
instrument  hard  disk  drive.  The  results  may  be 
opened  in  a  report  generation  program  like  MS 
Word  or  Excel  and  sent  to  a  local  or  network 
printer.    Once  the  results  have  been  reviewed 
and accepted by the operator the results may be 
sent  directly  to  a  network  folder  for  processing 
and direct entry into LIMS.  

 

Once  analysis  is  complete  the  instrument 
actuates the rapid cool-down system.  The cool-
down  system  is  capable  of  reducing  the 
temperature  of  the  sample  from  300  to  30°C  in 
about  8  minutes  in  a  laboratory  with  ambient 
conditions  at  20°C.  User  can  also  manually 
switch  to  standard  air  cooling.  Cooling  the 
system will take a longer period of time however 
fan noise will be reduced.  
 

1.4 

Instrument Specifications 

 

Models: 

K71000 

Electrical 

Requirements: 

 
115/230VAC 50/60Hz 
User selectable via switch 
located at the back of the 
analyzer 

Temperature 

Measuring Range: 

 
Ambient to 405°C (761°F)  

Flash Detection 

System: 

 
Differential 
Thermocouple,  
Ionization Detection 

Stirring Speed: 

0 to 300 RPM 

Heating Rate:  

In accordance to ASTM 
D93 procedure A, B, C  
Quick Test 

(≈12°C/min) 

Ignition Source:  

Gas and Electric Igniter 
both equipped on cover 
assembly: 
- Software Selectable 
- User-Friendly Manual 
Switching 

Mechanical Lift: 

Three (3) position 
mechanical lift system of 
cover / motor assembly, 
software selectable: 
Open 

– Clean – Test 

Cleaning: 

Cover / Motor assembly 
lift features convenient 
cleaning position. At this 
position the user can 
easily remove the cover 
assembly and clean with 
appropriate solvent 
between tests. 

Ignition Test 

Frequency: 

 
User selectable on per 
method basis 

Cooling: 

Integrated Dual Fan 
System: First (1) directly 
to cup, Second (2) to cool 
environment around test 
cup  

Barometric 

Pressure 

Correction: 

 
 
Automatic Barometric 
Pressure Transducer for 
Flash Temperature 
Correction 

Interface Port: 

RS232:1 
USB:2 (1

–Front, 1–Back) 

Ethernet:1 (LIMS) 

Gas Supply / 

Connection: 

 
Ignition Gas: LP Gas 
(Propane / Butane) or 
Natural Gas.  
Min. 3.45kPa (0.5psi). 
Max. 6.90kPa (1.0psi) 

Display: 

8.4" LCD Touch Screen 

Temperature 

Sensor: 

 
PT-100 in S/S sheath,  

User 

Programmable 

Methods & Data 

Storage: 

 
 
 
Capable of storing 
unlimited data runs and 
test methods 

Содержание K71000

Страница 1: ...enue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Страница 2: ......

Страница 3: ...e statement is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www dpa...

Страница 4: ...As a minimum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compact fl...

Страница 5: ...Safety Considerations 8 2 2 Unit Design 8 2 3 Contaminated Samples 9 2 4 Safe Quick Test Method 9 2 5 Fire Detection Suppression and Draft Shield Dome System 9 2 6 Equipment Modification and Replacem...

Страница 6: ...ility for losses arising from the negligent or improper use of our equipment nor for losses arising from the improper use or use other than intended Only deliberate action or gross negligence are grou...

Страница 7: ...ng needs Our products are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument service to better understand your r...

Страница 8: ...The scope of the method covers the analysis of products in the temperature range of 40 to 360 C but this automated apparatus extends the range from dew point 10 C to 400 C Internet Link to Summary AST...

Страница 9: ...ange Ambient to 405 C 761 F Flash Detection System Differential Thermocouple Ionization Detection Stirring Speed 0 to 300 RPM Heating Rate In accordance to ASTM D93 procedure A B C Quick Test 12 C min...

Страница 10: ...fety problems associated with the use of this equipment It is the responsibility of any user of this equipment to investigate research and establish appropriate safety and health practices and determi...

Страница 11: ...prescribed end point prior to running a quick test The set maximum run temperature is solely at the user s discretion We recommend this method of screening for ALL samples 2 5 Fire Detection Suppress...

Страница 12: ...emoved WARNING Electrical hazards exist inside the instrument case Lethal voltages and current are present Only factory authorized or otherwise qualified personnel may repair or service this instrumen...

Страница 13: ...r claim if requested When submitting a claim for shipping damage request that the carrier inspect the shipping container and equipment Do not return goods to Koehler without written authorization 3 3...

Страница 14: ...results outside the above mentioned criteria Ventilation A fume hood or exhaust system is required after performing a test Flammable vapors and or steam are generated during operation and must not be...

Страница 15: ...K71000 Automatic PMCC Flash Point Analyzer Operation and Instruction Manual K71000 Manual 13 4 Descriptions 4 1 Front of Instrument Figure 1 Instrument Descriptions_Front...

Страница 16: ...load factory software updates 4 Touch Screen Control Panel Main control point of instrument Please refer to section _ _ for full operational details 5 Leveling Feet To make adjustments to instrument l...

Страница 17: ...K71000 Automatic PMCC Flash Point Analyzer Operation and Instruction Manual K71000 Manual 15 4 2 Back of Instrument Figure 2 Instrument Descriptions_Back...

Страница 18: ...reen control panel 15 USB Port For connection to Keyboard or Mouse 16 RS232 Port For connection to external PC 17 Ethernet Port For connection to LIMS system 18 CPU Port For programming of internal CP...

Страница 19: ...etection thermocouple will automatically detect this and initiate re lighting of the flame 27 Pilot Flame The pilot flame ensures the test flame is lit prior to each test dip 28 Electric Ignitor Alter...

Страница 20: ...orities Key The fourth or factory level is designed for administrative purposes only and specifically allows for access to the factory settings menu From this screen the user can also Change User Pass...

Страница 21: ...t the which operations they want a particular user level to have access to The user can choose to turn on or off access to the Diagnostics Calibration and Custom Test Menus Click Save when complete Fi...

Страница 22: ...eset all global parameter settings back to the factory default Figure 7 Last Global Parameter Pop Up 5 Main Home Screen After clicking Yes or No to the Last Global Parameter pop up the user will be br...

Страница 23: ...d to a Laboratory LIMS system by clicking this button Print This function is only accessible when the user is in the results database screen Data can be sent to a printer by clicking this button Save...

Страница 24: ...dard Test Parameters Change Save minimum temperature in which the instrument lift can open safely and the temperature in which the cooling fan can be turned off Inst ID Setup Change Save Instrument ID...

Страница 25: ...Graph Tools The Graph screen features the following functions a Zoom In b Zoom Out c Pan XY Provides cursor that can be moved along graph and display data d Restore Resets X and Y Axis to default posi...

Страница 26: ...Dip Home Brings flame dip mechanism to home position b Find Lift Home Brings mechanical lift to Home Position also considered test position c Clean Position Brings mechanical lift to Clean Position h...

Страница 27: ...med Two Point Calibration RTD Correction Table User CRM Calibration Two Point Calibration as shown in Figure 14 below is performed using an external PT100 Simulator Box Available as an Additional Acce...

Страница 28: ...in which the user can input offset values for any of temperature values shown This table is particularly convenient for those with Certified Digital Contact Thermometers or Certified Liquid in Glass...

Страница 29: ...Material CRM samples using a Gas Ignition source 1 Input the certified flash point value from the CRM bottle label or test certificate into the first column titled CRM Flash Value 2 Perform a series...

Страница 30: ...erial CRM samples using the Electrical Ignition Source 1 Input the certified flash point value from the CRM bottle label or test certificate into the first column titled CRM Flash Value 2 Perform a se...

Страница 31: ...ric Pressure Correction Feature As pictured in Figure 18 below the user can Enter Atmospheric Pressure in Kilopascals kPa into the empty field The Measured Barometric Pressure is determined by the ana...

Страница 32: ...C value as pictured in Figure 19 below This method of calibration is useful when a material of known flash point is tested and the analyzer is displaying a value greater than the known value e g know...

Страница 33: ...RPM Heating Rate C min End Test Temperature EFP X C Fan Off Temperature C Pre Heat Limit Temperature C Pre Heat Ramp Rate C min Open Lift Temperature C Soak Temperature C Typically used only for Aspha...

Страница 34: ...parameter data will automatically be sent to the database Each test is listed in a single row with the information separated by column The user also has the option to delete or update the database lis...

Страница 35: ...and clicking on the Print button will cause a temporary Print Result Screen to appear followed by a Printer Options Menu See Figures 22 and 23 below Once the appropriate printer is selected click the...

Страница 36: ...n the user first enters this menu the display screen will be blank Clicking on the Monitor Traffic box will begin displaying the data communication At this time a screenshot can be taken of the curren...

Страница 37: ...for a Test Dip The Dip On Off Switch will activate or deactivate the dipping mechanism b Adjust Lift Mechanism Pressing the Up and Down Arrows will adjust the level at which the Lift Mechanism will b...

Страница 38: ...d See Figure 26 below The corrected flash point value is prominently displayed at the center of the screen Additionally the Type of Ignition used to perform the test is displayed as well as the Sensor...

Страница 39: ...authorize the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product i...

Страница 40: ...K71000 Automatic PMCC Flash Point Analyzer Operation and Instruction Manual K71000 Manual 38 Notes...

Страница 41: ...K71000 Automatic PMCC Flash Point Analyzer Operation and Instruction Manual K71000 Manual 39 Notes...

Страница 42: ...K71000 Automatic PMCC Flash Point Analyzer Operation and Instruction Manual K71000 Manual 40 Notes...

Отзывы: