K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual
-
10-
any fumes that may be emitted during the test.
Conduct tests behind a safety shield, especially
if you are not sure of the flash point or the
ignition temperature. Wear appropriate personal
protective equipment (PPE) during testing.
NOTE:
ASTM D93 specifically notes that
a hood should not be operated during testing as
excessive drafts can cause erroneous results!
Exercise caution in removing and replacing the
cup in the furnace. Hot surfaces and substances
may cause injury if they come in contact with the
operator.
All connections should be properly made and
tested for leaks prior to carrying out the tests.
The ignition gas source should be equipped with
a proper safety regulator capable of delivering a
maximum of 5 psi of pressure to the test flame
assembly. Some propane regulators give
pressure in inches of water (inH
2
0) at 60°F. A
standard 11 inches of water (inH
2
0) regulator
should give sufficient gas for normal use.
Minimum pressure required is about 7 kPa or
1.0 psi.
The flame suppression (inert gas) source should
be regulated to between 20 and 30 psi for safe
use.
2.9
Other Important Hazards
Working surfaces on the instrument may
become hot enough to produce severe burns
even with momentary contact.
Open flames and electrically heated resistance
elements are present. Direct momentary
contact can cause severe burns.
The draft shield around the sample area also
functions as a mechanical guard. The
instrument should never be used if this guard is
removed.
WARNING:
Electrical hazards exist inside
the instrument case. Lethal voltages and
current are present. Only factory authorized or
otherwise qualified personnel may repair or
service this instrument. Only original parts may
be used for repair. No substitution parts or
alterations to this equipment are allowed; doing
so may void the warranty.
PC Motherboard
WARNINGS
- Due to high voltages inside the instrument,
under no circumstances should the instrument
be opened or access covers removed while
energized.
- Lethal voltages are present inside power
supplies. Under no circumstances should
Power Supply covers be removed. Power
Supply warranties are voided once the cover
is removed.
- For indoor use only. Please operate the
power supply in a safe and dry environment.
- Do not insert any objects into the open
ventilation or fan grill area of the power supply.
3 Getting Started
The instructions for preparing the equipment
assume that the user is aware of the contents of
this document, which lists the warranty
conditions and important precautions.
3.1
Packing List
K71000 Automatic PMCC Flash Point
Analyzer, 115/230V 50/60Hz
Brass Test Cup with Handle
Combination Gas/Electric Igniter
Stirrer Shaft and Blade Assembly
PT-100 Temperature Probe / Thermal
Detector
Ionization Detector
Cover Assembly
Draft Shield
Stylus
Temperature Calibration Cable
Power Cords, 115V & 230V
K71000-Manual K71000 Automatic PMCC
Flash Point Analyzer Operation and
Instruction Manual
Installed in the K71000
Panel PC and Microprocessor Controllers
Operating Software, Instruction Manual
PDF, Recovery Software on Hard Disk Drive
Thermocouple and Ionization Detection
Fire Sensing and Suppression System
Collision / Interference Detection
Mechanical Lift System
Dual Fan Cooling System