![Koehler K71000 Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/koehler/k71000/k71000_operation-and-instruction-manual_1984929030.webp)
K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual
-
28-
User CRM Calibration_Electric CRM Calibration
as shown in
Figure 17
below is a user
defined table that allows for the input of flash point values of up to four (4) Certified Reference
Material (CRM) samples using the Electrical Ignition Source.
1.
Input the certified flash point value from the CRM bottle label or test certificate into the first
column titled
CRM Flash Value
2.
Perform a series of four flash point tests with the analyzer using the CRM sample
3.
Input the Flash Point Value as determined by the Analyzer into each
Run Column 1 - 4
4.
Click the
Save Changes
button to confirm the instrument calibration
Figure 17: Calibration Screen_Electric CRM Calibration