79
Time and Date
Control of the calibration and cleaning cycles is based on the time and
date of the integrated real-time clock.
In measuring mode the time is shown in the lower display.
When using InduCon sensors, the calibration data is written in the
sensor head.
In addition, the logbook entries (cf Diagnostics) are provided with a
time stamp.
Please note:
After prolonged power outage (> 5 days) the time display is
•
replaced by dashes and cannot be used for processing.
Enter the correct time.
There is no automatic switchover from winter to summer time!
•
Be sure to manually adjust the time!
Confi guration
Menu item
Action
Choices
Tag number
Select character using
keys,
select next digit using
keys.
Confirm with
enter
A...Z, 0...9, – + < > ? / @
The first 10 characters are
seen in the display with-
out scrolling.
Tag Number (“ TAG“)
You can enter a designation for the point of measurement (tag
number) in the lower display line. Up to 32 digits are possible.
Pressing
meas
(repeatedly) in the measuring mode indicates the tag
number.
Being part of the device confi guration, the “TAG“ can be read out via
IrDA.
A standardized tag number helps, for example, to correctly re-install a
device after repair.