55
Confi guration
Menu item
Action
Choices
Output current
during error message
Select ON or OFF using
keys.
Confirm with
enter
ON/
OFF
Output current
during HOLD
LAST: During HOLD the
last measured value is
maintained at the output.
FIX: During HOLD a value
(to be entered) is main-
tained at the output.
Select using
Confirm with
enter
LAST
/FIX
Output current for
HOLD FIX
Only with FIX selected:
Enter current which is to
flow at the output during
HOLD
Enter value using
keys.
Confirm with
enter
00.00...22.00 mA
(
21.00 mA
)
HOLD active
21
4
HOLD active
Output current
[mA]
Output signal HOLD
Setting FIX = 21.0 mA
Output signal for HOLD
LAST setting
Output Signal During HOLD:
5