1. Description
2
1.1
Unit Functions 2
1.2
Arrangement in
Vacuum Systems
2
1.3
Mechanical Attachment
2
1.4 Electrical
Connection
2
1.5 Pneumatic
Connection
2
1.6
Switching on/off
2
1.7
Explanation of Display
Elements
3
1.8
Explanation of Control
Panel
3
2.
General Safety
Precautions
3
3. General
Operating
Instructions
3
3.1
Operating Conditions
3
3.2 Starting
3
3.3
Shutdown
4
4.
Manual Operation
4
4.1
Controlling Profiles
4
4.2
Setting Values
5
4.3
Running
5
4.3.1
Controlling
5
4.3.2
Controlling with soft
setpoint start
5
4.3.3
Controlled Pressure
Reduction
5
4.4
Temporary Mode
6
4.4.1
Function and Operation
6
4.4.2 Aeration
during Distillation 6
4.4.3
Experimental Use
6
4.5
Uncontrolled Evacuation/
Drying Mode
6
5.
Automatic Operation
7
5.1
Controlling Profiles
7
5.2
Setting Values
7
5.3
Running
7
6.
Changing of Pressure unit 8
7.
Calibration
8
8.
Trouble Shooting
8
9. Specifications
8
10.
Ordering Information
8
10.1
Accessories
8
10.2
Vacuum Controller and
Vacuum Valves
8
1. Description
NC 800: For use at LABOPORT vacuum
systems.
NBC 800: Workstation vacuum control-
ler for use e. g. at vacuum supply
points.
Technical data: see chapter 8.
1.1 Unit Functions
The vacuum controller NBC 800 regula-
tes vacuum on an individual basis in
the laboratory working place. Several
different functions are possible:
Evacuation at pressure setpoint;
regulation by pressure setpoint and
pressure differential (manual ope-
ration);
defined post distillation;
experimental use;
automatic search for distillation
point (automatic operation).
1.2 Arrangement in Vacuum Systems
See fig. 1.
1.3 Mechanical Attachment
The Vacuum Controller NBC 800 can be
attached to a round support by means
of the connection on its back side
(maximum support diameter: 13 mm).
Other types of connection on request.
Attachment in LABOPORT® Sys-
tems: see LABOPORT Operating
Manual.
1.4 Electrical Connection
Electrical supply:
The socket is located on the back
of the unit.
Cooling water valve and vacuum
valve:
Both can be connected either to
the side socket or the socket on
the back (6-pin). Proper connec-
tions are ensured by means of
different pin arrangements.
Insert plug into a properly installed
safety socket.
Electrical Connection in
LABOPORT® Systems: see
LABOPORT Operating Manual.
1.5 Pneumatic Connection
The pneumatic connection is made
at the valve block on the hose shaft
(tube ID 10mm).
Pneumatic Connection in
LABOPORT® Systems: see
LABOPORT Operating Manual.
1.6 Switching on/off
The vacuum controller is switched
on and off by the switch on the
side.
Symbols:
= Position in the illustration,
= Important point,
= Task,
= Advice to users,
= Product information
Warning
2
Translation of original Operating Instructions, english, KNF 121205-121650 02/10
Fig. 1: Arrangement of vacuum controller in vacuum system
Specification:
1
Vacuum controller NBC 800
2
Vacuum pump
3
High Performance Condenser of
Vacuum System
4
Vacuum valve
5
Venting valve (internal)
6
Condensate recipient
7
Vacuum installation
8
Suction line
9
Cleaning gas connection/Aeration
10 Coolant valve (accessory)
11 Cooling water tubing