Installation, mounting and connection
LABOPORT Pumps and Systems
24
Translation of original Operating Instructions, english, KNF 121211-121345 01/10
7.2. Baseplate
Pump disconnected from mains
1. Unscrew the pump`s rubber feet (
1
) anti-clockwise.
2. Install the mounts (
2
) onto the pump.
3. Stand the pump (fitted with the mounts) onto the baseplate
4. Aligne the mounts on the pump and the baseplate.
5. Slide the locating pin (see fig. 2/
13
) through the mount holes
on the baseplate and on the pump until they are fully engaged.
6. Check that the pump is securely fastened to the baseplate via
both mounts.
7.3. Separator
Baseplate mounted (see chapter 7.2)
Pump disconnected from mains
Qty Material
1
Phillips-head screwdriver No. 3
Tab. 14
1. Slide the container for the separator (see fig. 6/
4
) into the
baseplate’s upward-pointing slotted bar.
2. Fix the holders by tighten the screw in the bottom of them.
3. Place the glass separators into the holders.
Laying hoses: depending on system configuration see chapter
7.7.1 (system SR), 7.7.2 (system SH), 7.7.3 (system SC), 7.7.4
(system SCC) or 7.7.5 (retrofitting from system SC to system
SCC).
Condition
Fig. 9: Installation of the mounts at
the pump
Conditions
Tools and material