7. Installation, mounting and connection
Only install and operate the pumps/systems under the operating
parameters and conditions described in chapter 5, Technical data.
Observe the safety precautions (see chapter 3, page 6).
Before installation, store the pump/the accessories at the
installation location to bring it up to room temperature.
See chapter 5, Technical data, for the dimensions of system.
Install the pump/system so that the motor fan can intake suffi-
cient cooling air.
Make sure that the installation location is dry and the
pump/system is protected against rain, splash, hose and drip
water.
Choose a safe location (flat surface) for the pump/system.
Protect the pump/system from dust.
Protect the pump/system from vibrations and jolts.
7.1. Connect pump
Only connect components to the pump which are designed for
the pneumatic data of the pump (see chapter 5.1, page 10).
If the pump is used as a vacuum pump, safely discharge the
pump exhaust at the pump’s pneumatic outlet.
A marking on the pump head shows the direction of flow.
1. Remove the protective plugs from the pneumatic connectors of
the pump.
2. Connect the suction line and pressure line.
3. Lay the suction and pressure line at a downward angle to
prevent condensate from running into the pump.
4. Insert the power cable’s plug into a properly installed
shockproof socket.
Dimensions
Cooling air supply
Installation location
Connected components
Pump exhaust