LABOPORT Pumps and Systems
Troubleshooting
Translation of original Operating Instructions, english, KNF 121211-121345 01/10
45
10.3. System with two Vacuum Controllers
See also chapter 10.2, system with one vacuum controller.
System does not work, despite one Vacuum Controller (both controllers) being set on
"Regulation Mode"
Cause
Fault remedy
In the electrical supply unit the
connecting wires for the vacuum
controllers are plugged in incor-
rectly.
Exchange the plugs between vacuum controller 1 and vac-
uum controller 2.
In the electrical supply unit the
connecting wires for the vacuum
valve are plugged incorrectly.
Exchange the plugs between vacuum valve 1 and vacuum
valve 2.
Tab. 35
10.4. Fault cannot be rectified
If you are unable to determine any of the specified causes, send
the pump to KNF Customer Service (see page 47 for the address).
1. Flush the pump to free the pump head of dangerous or ag-
gressive gases (see chapter 9.2.1, page 37).
2. Remove
the
pump.
3. Clean the pump (see chapter 9.2.2, page 37).
4. Send the pump to KNF with a filled out decontamination decla-
ration (see chapter 12) and specification of the medium trans-
ferred.