Servicing
LABOPORT Pumps and Systems
38
Translation of original Operating Instructions, english, KNF 121211-121345 01/10
9.3. Changing Diaphragm and Valve
Plates/Sealings
Pump is switched off and mains plug is removed from the
socket
Pump is clean and free of hazardous materials
Tubes removed from pump’s pneumatic inlet and outlet
Pump is removed from baseplate (if pump is integrated into a
system)
Qty Material
1
Phillips-head screwdriver No. 2
1
Service Set (see chapter 11.1, page 46)
1 Felt-tip
pen
Tab. 25
Always replace diaphragm and valve plates/sealings together
to maintain the pump performance.
With multi-head pumps, parts of the individual pump heads can be
confused.
Replace the diaphragm and valve plates/sealings of the indi-
vidual pump heads consecutively.
WARNING
Health hazard due to dangerous substances
in the pump!
Depending on the substance transferred, caustic
burns or poisoning are possible.
Wear protective clothing if necessary, e.g.
protective gloves.
Flush pump before replacing the diaphragm and
valve plates/sealings (see chapter 9.2.1, page
37).
Conditions
Tools and material
Information on procedure