41
AZURA® Detector UVD 2.1S Instructions, V6820
9. Transport and storage
Regarding the following information, carefully prepare the device for
transport or storage.
9.1 Taking the device out of operation
The device is switched off.
Procedure
Process
1.
Pull the power plug out of the socket and then out of the device.
2.
Pack the power cable together with the device.
Disconnect all electrical connections. Remove the accessories and pack
the device for transport or storage.
9.2 Packing the device
Original packaging: Ideally you should use the original transport
packaging.
Lifting: Grab the device around the center of both sides and lift it into
the packaging. Do not hold onto front cover or leak tray, as these parts
are loosely attached to the device.
9.3 Transporting the device
Documents: If you want to return your device to KNAUER for repairs,
enclose the „
Service request form and decontamination report
“ which
can be downloaded from our website.
Device data: For a secure transport, note the weight and dimensions of
the device (see chapter 11 on page 43).
9.4 Storing the device
Flushing solution: Pay attention that all hoses and capillaries have
been emptied or filled with flushing solution (e. g. isopropanol) before
storage. To prevent algae formation, do not use pure water.
Seals: Close all inputs and outputs with cap fittings.
Ambient conditions: The device can be stored under ambient condi-
tions which are listed in the Technical Data section (see chapter 11 on
Prerequisites
Next steps