11. Technical data
11.1 Detection
Detector type
Small variable single wavelength UV/VIS
detector
Small variable single wavelength UV/VIS
detector with fiber optic connectors
Detection channels
1
Light source
Deuterium (D
2
) lamp with integrated GLP chip
Wavelength range
190–500 nm
Spectral bandwidth
13 nm at H
α
line (FWHM)
Wavelength accuracy
± 3 nm
Wavelength precision 0.7 nm (ASTM E275-93)
Noise
± 2.0 × 10
-5
AU at 254 nm (ASTM E1657-98)
Fiber optics version:
± 2.5 × 10
-5
AU at 254 nm (ASTM E1657-98)
Drift
3.0 × 10
-4
AU/h at 254 nm (ASTM E1657-98)
Fiber optics version:
4.0 × 10
-4
AU/h at 254 nm (ASTM E1657-98)
Linearity
> 2.0 AU at 270 nm (ASTM E1657-98)
Time constraints
0.0 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 s
Integration time
Automatic
11.2 Communication
Max. data rate
50 Hz (LAN), 20 Hz (analog), 10 Hz (RS-232)
Interfaces
LAN (RJ-45)
RS-232 (SUB-D 9, service only)
Pin header
Analog (RCA cinch connector)
Control
Manually via front panel
Mobile Control
Chromatography software
Event control
Analog: wavelength control
Terminal protocol
43
AZURA® Detector UVD 2.1S Instructions, V6820