Troubleshooting
AZURA® Pump P 2.1L Instructions V6840
Message
Solution
Cannot initialize LAN
Check cables and connections in local
area net work.
Cannot operate with an
empty link
Create a link.
Cannot purge during the
run!
End method and start purging.
Cannot read data from
FRAM
Switch the device off and on. Inform the
Techni cal Support of the manufacturer in
case the sys tem message repeats itself.
Cannot read RTC
Switch the device off and on. Inform the
Techni cal Support of the manufacturer in
case the sys tem message repeats itself.
Cannot start time table
Edit the data in the chro matography
software.
Cannot use non-existing
compo nent!
Change the setup settings or change the
gradi ent in the program or in setup.
Cannot write data on
FRAM
Switch the device off and on. Inform the
Techni cal Support of the manufacturer in
case the sys tem message repeats itself.
Component set tings not
compati ble with gradient
setup!
Change the setup settings or change the
gradi ent in the program or in setup.
E
Error input acti vated
Device error, change device settings.
G
GUI communica tion
failed
Switch the device off and on. Inform the
Techni cal Support of the manufacturer in
case the sys tem message repeats itself.
I
Instrument in stand alone
mode
Change the entry in the Setup menu.
Change the entry in the chromatogra-
phy soft ware.
Instrument remote
controlled
This entry is not executable. Quit
software.
Invalid index in time
table
Change the entry in the program line.
Invalid line number
Change the entry in the program line.
Invalid time in time table
Correct the time entry.
L
Leak sensor not present
Switch the device off and then on.
If the leak sensor is still not present,
contact the Technical Support of the
manufacturer.
Leak was detected
Switch off the device. Remove the leak
and start the device afterwards.
Line in time table is
empty
Edit the program line.