60
Transport and storage
AZURA® Pump P 2.1L Instructions V6840
9.4.1
Storing the pump head
If you want to store pump heads separately, proceed as follows:
The pump head has been dismounted (see chapter “8.3.2 Dismounting
Syringe
Flushing solution
Process
Procedure
1.
Fill the flushing solution into a syringe and inject into the capillary at
the pump head inlet. Leave it for 5 minutes.
2.
Flush with suitable flushing solution.
3.
Fill the pump head with isopropanol.
4.
Seal the inlets and outles with cap fittings.
The pump head can be stored under the ambient conditions described in
chapter “11. Technical data” on page 62.
Device defect
Residuals chemicals can damage the pump head if they are not being
removed before storage.
For reuse, fl ush the pump head and fi ll it with isopropanol.
Before storage, seal the inlets and outlets with cap fi ttings.
Prerequisite
Materials
Storage conditions