AZURA® Pump P 2.1L Instructions V6840
30
Installation and initial startup
4.10.4 Firmware Wizard: Setting a static IP address
Note:
More information about LAN settings can be found in the
Mobile Control Software Instructions in the chapter "Firmware Wizard”
(document no. V6851-2).
1
Text box for serial
number of the
device
2
Setting IP address
manually
3
Text box for IP
address
4
Text box for subnet
mask & gateway
5
Confirm changes
Fig. 29:
Network settings in Firmware Wizard:
The device is switched on.
Firmware Wizard is installed and running.
Process
1.
In Firmware Wizard, click <Reset LAN Settings...>.
2.
The window <Device connection settings> opens. Enter serial num-
ber of the device into the text field <Target device serial number>
1
.
3.
Select option <Use the following IP address>
2
.
4.
Enter the IP address into the text field <IP address>
3
.
5.
Optionally, adjust subnet mask and gateway
4
.
6.
Click <Reset Conn. Settings>
5
to accept changes.
7.
Restart the device (recommended).
The device is now accessible via the static IP address.
4.10.5 Firmware Wizard: Setting a dynamic IP address
The device is switched on.
Firmware Wizard is installed and running.
1
2
3
4
5
Prerequisites
Procedure
Result
Prerequisites