AZURA® Pump P 2.1L Instructions V6840
Installation and initial startup
27
Process
1.
Make sure that there is no overlap between the IP addresses of the
router and the corporate network.
2.
In case of an overlap, change the IP address range of the router.
3.
Use the patch cable to connect the router WAN port to the compa-
ny network.
4.
Restart all devices, including the computer.
The LAN is now integrated into the company network.
4.9.6
Controlling several systems separately in LAN
Devices connected to a LAN communicate through ports, which are part
of the IP address. If more than one chromatography system is connected
to the same LAN and you plan on controlling them separately, you can
use different ports to avoid interference. Therefore, the port number for
each device must be changed and this same number must be entered
into the device configuration of the chromatography software. We recom-
mend to use the same port number for all devices in the same system.
Note:
The port is set to 10001 at the factory. You must use the same num-
bers in the device configuration of the chromatography software as in the
device, otherwise the connection fails.
Process
1.
Find out port number and change it on the device.
2.
Enter the port number in the chromatography software.
The connection is established.
4.10 Setting IP addresses via software
Note:
Check the IT security standards for your lab before intervening in
the LAN settings.
Static IP addresses are required to run certain chromatography software,
e.g. Purity Chrom®. For a comprehensive overview on how to set static IP
addresses for PurityChrom®, refer to the document „PurityChrom® Instal-
lation Guide“ on the PurityChrom® installation CD.
For Mobile Control and Firmware Wizard, it is possible to set a fixed (sta-
tic) or dynamic (DHCP) IP address via software.
4.10.1 Mobile Control: Setting a static IP address
Note:
The device is preset to a dynamic address (DHCP). To ensure a
constant LAN connection between the chromatography software and
the instrument, we recommend switching the instrument to a static IP
address for certain applications. More information about LAN settings
Mobile Control Software Instructions
(document no.
V6851-2).
Procedure
Result
Procedure
Result
PurityChrom®