C-2
APPENDIX C Compass Calibration
HydroChart 3500 Sonar System Operations and Maintenance Manual
P/N 11210087, Rev. 03
C.2 Setting up for the Compass Calibration
When setting up for the compass calibration, refer to Figure C-1 on page C-1. In
addition, the cable harness is fabricated from a SubConn 5-pin bulkhead connector
(P/N MCBH5F-SS) and an OceanServer digital compass cable (P/N 19-00061-24)
which includes the DB9 connector and the battery pack. To fabricate the cable
harness, cut off the small connector on the compass cable and splice the cut wires
to the wires of the 5-pin bulkhead connector in accordance with Table C-1
.
To set up for the compass calibration:
1.
Connect the DB9 connector of the cable harness to the laptop PC and the 5-pin
bulkhead connector to the cable of the compass assembly.
2.
Place the laptop PC, the cable harness with the battery pack, and the compass on
the table or cart.
3.
Secure the compass with tape. It should be as level as possible.
C.3 Performing the Compass Calibration
To perform the compass calibration:
1.
Turn on the laptop PC and start the OS5000 Windows Compass Demo
Program.
2.
Click
Configure
.
The Setup dialog box opens as shown in Figure C-2.
Table C-1:
Cable Harness 5-Pin Bulkhead Connector to DB9 and Battery
Connectors Pin-to-Pin Connections
5-PIN BULKHEAD
CONNECTOR PIN NO.
DB9 CONNECTOR PIN
NO.
BATTERY
TERMINAL
WIRE
COLOR
1
—
Pos
Red
2
3
—
Blue
3
2
Green
4
5
Neg
Black
5
—