KI 7010B UM-11
27
PERFORMANCE VERIFICATION TEST
Model:
Report No.:______________ Date:________________
Linearity / Accuracy Test
Wavelength =
Setting
Attenuator Setting, dB
Minimum Specification, dBm
(-0.3dB of Reference.)
Power Meter
Measurement results, dBm
Maximum Specification, dBm
(+0.3dB of Reference.)
1.
2.00
-2.30
-1.70
2.
5.00
-5.30
-4.70
3.
10.00
-10.30
-9.70
4.
20.00
-20.30
-19.70
5.
30.00
-30.30
-29.70
6.
40.00
-40.30
-39.70
7.
50.00
-50.30
-49.70
8.
60.00
-60.30
-59.70
Measurement Uncertainty: ___________dB
Table 4. Linearity/Accuracy Test Record for KI7010B Series