KI 7010B-UM-11
12
ABSOLUTE / RELATIVE DISPLAY MODE
Absolute/Relative display mode is a useful feature to allow user
switching between these two modes depending on which value is
required.
Absolute attenuation value is displayed in dB.
In relative mode, ‘REL
dB
’ shows on the top right side of the display.
Sometimes it is convenient to display attenuation relative to a
reference value, e.g.:
•
The display is zeroed at a particular point, and subsequent
readings are relative to this.
•
The display is set to an arbitrary value. For example, it can be
set to display the actual optical power coming out of the
attenuator when a light source is attached or to read total link
loss including the attenuator loss.
Setting up the attenuator to display actual power coming out
of it:
- connect a light source to the left port, and a power meter (for
referencing purpose only) to the right port of the attenuator.
Turn on all instruments and set them up with the same
wavelength.
- in Absolute display mode of the attenuator, set attenuation
level as required and press [ABS/REL] to switch to Relative
display mode.
“REL dB” will show on the top right of the
display.
- Press [-/+] to adjust the
attenuator’s
display to the value corresponds
to the
absolute
value displayed on power meter, then press [SET] to
save setting. Note that negative value settings on attenuator are
invalid.
At this point, the attenuator has been setup to display its output
power at the current set attenuation level.
- press [-/+] now to change attenuation level by the default 0.1dB
decrement/increment, will result in the attenuator displaying its
output power which corresponds to the new attenuation level set.
- press [Step] to enter Step mode to increment/decrement attenuation
level in the pre-set step value.
The step value can be reset by pressing [SET STEP] followed by [-
/+]. Press [SET] to save setting.
To return to the default 0.1 dB increment/decrement attenuation
adjustment mode, press [MENU].
-
To exit Relative display mode, press the [ABS/REL] button again or
[MENU].
Содержание KI 7010B Series
Страница 1: ...OPERATION MAINTENANCE GUIDE KI 7010B Series Automated Variable Optical Attenuator ...
Страница 5: ...KI 7010B UM 11 3 KI7000B SERIES AUTOMATED VARIABLE OPTICAL ATTENUATOR ...
Страница 32: ...KI 7010B UM 11 30 ...
Страница 34: ...KI 7010B UM 11 32 NOTES ...
Страница 35: ...KI 7010B UM 11 33 NOTES ...
Страница 36: ...KI 7010B UM 11 34 NOTES ...