![Kinemetrics EpiSensor FBA ES-T Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/kinemetrics/episensor-fba-es-t/episensor-fba-es-t_user-manual_1968295008.webp)
S
AFETY
2
E
PI
S
ENSOR
U
SER
G
UIDE
Specific Precautions
Follow the precautions below to ensure your personal safety and prevent
damage to the EpiSensor.
Power Source
The EpiSensor must be supplied with power either from a recorder or from
a customer-supplied
"
12V or
"
15V power supply (or a + 12V supply for
the single-supply option).
If you plan to power the EpiSensor from a recorder, connect the recorder to
a power supply/charger supplied by Kinemetrics, as described in each
recorder's user manual.
To supply power directly to the EpiSensor, you need a low-noise, regulated
"
12V or
"
15V power supply (or a + 12V supply for the single-supply
option) that is safely grounded and meets all applicable local regulations.
The EpiSensor will be damaged if the power is connected with the wrong
polarity.
User
-
Supplied Power
/
Charging System
If you supply your own power/charging system, be sure that the system
provides the correct voltage and current required by the EpiSensor under all
operating conditions. You are responsible for the safety of your charging
system. If you get power from the mains supply, be sure you have supplied
adequate grounding for all the equipment. If you supply your own batteries,
follow the manufacturer’s safety recommendations.
Sensor Grounding and Cabling
In some cases the EpiSensor will be a long distance from the recorder. In
these installations it is possible, due either to faulty AC wiring or extremely
high earth-return currents, for a high potential difference to exist between
the grounds at the two locations. When the cable is grounded at one end a
potentially lethal voltage can exist between the other end of the cable and
ground. Consider this danger during installation and get help from a
qualified electrician if this danger exists.
Do Not Operate in Explosive Atmosphere
The EpiSensor provides
no explosive protection
from static discharges or
arcing components.
Do not
operate the equipment in an atmosphere where
explosive gases are present.
!