61
Figure 33: Wrapping foil tape around outside of the connection
Slide the Shrink-N-Shield tubing forward, as shown in Figure 34, then up
over the plastic grommet ring.
Figure 34: Slipping Shrink-n-Shield tubing into place
Carefully
place the tubing so its cloth shield is up against-but not covering
over-the connector threads, as shown in Figure 35.
Figure 35: Shrink-n-Shield tubing in correct position (after shrinking)
Using even heat, shrink the tubing until it is tightly in place over the foil
tape (as shown in Figure 35), and begin the final assembly process:
1. Slide the connector shell forward and screw-tighten it onto the
connector head.
2. Slide the connector strain relief grommet forward.
3. Check the resistance from the connector shell to the shields at the
other end of the cable. The resistance will vary depending on cable
length, but should be very low (<0.3
%
/ 100').
4. Place the two halves of the cable clamp in place, then tighten its
screws.
5. Slide the Thermofit tubing into place and then shrink it