TOS8030
General 13
High-voltage test probes
The HP01A-TOS and HP02A-TOS are test voltage output
probes designed to be connected to our withstanding volt-
age tester.
The test probes are constructed so that a test voltage is
output only when the slide lever at the grip of a test probe
is held and the trigger is activated with one hand, and a
switch on the upper part of the probe is pressed with the
other; that is, operation requires two hands. Releasing
either hand outputs a STOP signal, shutting off the tester’s
test voltage.
This features are intended to prevent inadvertent output of
test voltages when using these probes.
• When using a test probe, do not connect it to the DUT while a test voltage
is being output from the probe. Also, do not disconnect the probe from the
DUT while a test voltage is being output from the probe.
Disconnecting the probe from the DUT while high voltage is being output
from the probe may result in damage to the DUT. Additionally, disconnect-
ing the probe from the DUT in the middle of testing may result in a residual
electrical charge in the DUT, posing a significant hazard.
For these reasons, the probe must be connected to the DUT before testing
begins. When ending the test, confirm that the LED on the probe is not lit,
then disconnect the probe from the DUT.
High-voltage test leads
Model number
Maximum usage voltage
Cable length
HP01A-TOS
4 kVac (rms) 50 Hz/60 Hz
5 kVdc
Approx. 1.8 m
HP02A-TOS
Approx. 3.5 m
Model number
Maximum usage voltage
Cable length
Remarks
TL01-TOS
5 kVac (rms) 50 Hz/60 Hz
5 kVdc
Approx. 1.5 m
Equivalent of TOS8030
accessory
TL02-TOS
Approx. 3.0 m
HP01A-TOS