16 Installation and Preparations for Use
TOS8030
■
Do not use the tester in locations where sensitive measuring
instruments or receivers are also being used.
Use of the tester may affect the reliability or accuracy of such instruments or
devices.
At a test voltage of more than 3 kV, corona discharges may occur that result in sig-
nificant wide-range RF emissions between the clips of the test leads. To minimize
these effects, keep the alligator clips as far apart as possible. Never allow an alliga-
tor clip or test lead to contact or approach the conductor surface (especially sharp
metal ends).
■
Provide sufficient space around the power plug.
Do not use excessive force to insert the power plug into an electrical outlet that
resists insertion/extraction. Do not install objects near the plug that make plug inser-
tion/extraction difficult.
2.3
Moving Precautions
When moving or transporting the tester to another installation site, observe the fol-
lowing precautions:
■
Turn the POWER switch off.
Moving the tester with the power turned on may result in electric shock or damage.
■
Disconnect all wiring.
Moving the tester with cables connected may result in breaks in the cables or may
cause the tester to fall, which could result in injury.
■
When transporting the tester, always use the dedicated pack-
ing materials.
Failure to use the dedicated packing materials may result in damage to the tester in
the event of a fall or due to vibrations during transport.
■
Be sure the operation manual be included.
2.4
Connecting the Power Cord
This product is designed to meet the standards for Overvoltage Category II (energy-
consuming equipment to be supplied from fixed installation) of the IEC standards.
Checking the supply voltage
Before connecting the power cord, check the tester’s supply voltage (100 V or 200 V).
The tester’s nominal input rating is indicated on the rear panel.