40 Remote Control
TOS8030
REMOTE connector pin configuration
•
The configuration of the REMOTE connector pin numbers are based on DIN
Standards. Please note that the pins are not arranged by pin number.
• If the REMOTE connector pin 3 is at level L, the tester switches from front panel
control to remote control.
Configure the control circuits so that pins 2 and 3 connect externally.
Control circuit example
Fig. 6-2
Example: Circuit 1
Fig. 6-3
Example: Circuit 2
Logic elements, transistors, photocouplers, or other elements may take the place of
contacts in Fig. 6-2. Fig. 6-3 shows one example.
Viewed
from the
panel face
Pin number
Signal name
Description
1
RSV
Leave this pin unconnected.
2
COM
Common terminal
3
ENABLE
Enables remote control at level L.
4
START
Testing begins at level L.
5
STOP
Stops testing at level L.
In this example, controlling
the START and STOP con-
tacts lets you operate the
tester in the same way as from
the front panel.
In this example, setting the
contact to the NO position
invokes a TEST state; return-
ing it to the NC position stops
testing.
1
2
3
4
5
NOTE
TOS8030
5
4
3
2
1
REMOTE connector
COM
START
STOP
TOS8030
5
4
3
2
1
COM
NO
NC
REMOTE connector