TOS8030
Panel Operations 31
■
Checking the firmware version
To check the firmware version, turn the POWER switch on pressing the STOP
switch.
After all the display units and indications light, the firmware version is displayed in
the voltmeter. The tester then changes to display the test conditions currently set.
Fig. 5-3
Firmware Version display
■
Do not turn the POWER switch on/off in quick succession.
After turning the POWER switch off, wait several seconds before turning it on
again.
Never perform a rapid on/off cycle of the POWER switch. The tester’s protection
function may be unable to protect the tester under these conditions, resulting in
electric shock or other problems.
Except during an emergency, never turn the POWER switch off while voltage is
being output.
5.2
Tester States and Indications
The front panel’s indicators indicate tester state.
Fig. 5-4
Indicators on the Front Panel
Example of indication of version 1.00
After all the display units and
indications light, the firmware
version is displayed in the
voltmeter.
When lit, this lamp
indicates that testing is in
(TEST state)
When lit, this lamp indicates
that the test result has been
determined to be PASS.
(PASS state)
When lit, this lamp indicates that
the protection circuit has been
activated.
(PROTECTION state)
If all four lamps are unlit, you
may press the START switch to
start testing immediately.
(READY state)
When lit, this lamp indicates
that the test result has been
determined to be FAIL.
(FAIL state)
progress.