TOS5051A/5050A
7-7
Chap
.7
RS-232C Interf
ace
7.6
RS-232C Commands and Responses
This section describes the commands that the tester supports one by one.
•
Commands are not case-sensitive. Responses are returned in upper-case.
•
The response for command error is ERROR.
•
The command inside the parentheses is the abbreviated form of the command.
*IDN?
Queries the model name and ROM version of the tester.
Response
Returned as in the example below.
Fig. 7-2
Response example of *IDN?
*RST
Resets the tester (the talk mode is set to default (0) in this version).
Response
Returns OK.
• The talk mode setting speci
fi
ed by a command is reset to the panel setting when
the tester is power cycled.
STATUS? (STAT?)
Queries the tester status.
Response
Returns any of the following:
PROTECTION, TEST, READY, PASS, U_FAIL, L_FAIL, or ELSE
(ELSE indicates a status other than PROTECTION, TEST, READY, PASS,
U_FAIL, or L_FAIL.)
KIKUSUI ELECTRONICS CORP.,TOS5051A,0,1.00