6-10
TOS5051A/5050A
The following actions will take place.
♦
The tester terminates the test and cuts off the test voltage.
♦
The “TEST” message disappears.
♦
The TEST signal of the SIGNAL I/O connector is turned OFF.
♦
The DANGER lamp turns OFF. (It does not go out as long as a residual test
voltage remains in the output circuit.)
♦
The H.V ON signal of the SIGNAL I/O connector is turned OFF. (It is not
turned off as long as a residual test voltage remains in the output circuit.)
♦
The “UPPER FAIL” or “LOWER FAIL” message appears.
♦
Turn ON the U FAIL signal and L FAIL signal of the SIGNAL I/O connector.
♦
The buzzer sounds.
The FAIL judgment does not reset automatically. To reset it, press the STOP switch.
Repeating the test
Within approximately 0.2 seconds after the test was successfully terminated with a
PASS judgment, the tester will automatically reset to the READY status, allowing
you to repeat the test simply by pressing the START switch. The same applies also
when the tester is controlled remotely.
• The tester will not automatically reset to the READY status when in any of the
following cases:
♦
When the PASS HOLD switch of the TEST MODE switches on the rear panel has
been set to ON. (For details, refer to section 6.5, “Settings for Special Test
Modes” on Page 6-29 .)
♦
As long as the cause of the PROTECTION status remains. (For details, refer to
[32] PROTECTION message in section 4.2, “Display Items” on Page 4-18 .)
♦
When the STOP switch is pressed (including the STOP switch of the remote con-
trol box).
♦
When the START switch remains in the pressed state (including the START
switch of the remote control box).
When the test is terminated with a FAIL judgment, press the STOP switch to reset
the tester to the READY status and then press the START switch. The same
applies also when the tester is controlled remotely.
NOTE