4-6
TOS5051A/5050A
[4]
REMOTE connector
This connector is used to remotely control the start/stop of the test operation from
an optional device such as remote control box (RC01-TOS or RC02-TOS) or HV
test probe (HP01A-TOS or HP02A-TOS).
When you connect the plug of the optional device, the remote control circuit is
established, the “REMOTE” message appears, and the local START switch is dis-
abled. In this case, the tester enters the PROTECTION status (“PROTECTION” is
lit) and the high voltage output is cut off. The STOP operation can be done locally
from the STOP switch on the front panel or remotely from the optional device.
When you disconnect the plug of the optional device, the “REMOTE” message dis-
appears and the START switch is enabled. The tester enters the PROTECTION sta-
tus (“PROTECTION” is lit) and the high voltage output is cut off.
• Use the optional HV test probe (HP01A-TOS or HP02A-TOS) at a test volt-
age of 4 kV or less in AC test mode and 5 kV or less in DC test mode.
• When controlling the start/stop function remotely using a device other than one of
the optional devices of the tester, do not use the REMOTE CONTROL connector
on the front panel. Use the SIGNAL I/O connector on the rear panel, instead.
• Do not remote-control the tester simultaneously from both the REMOTE CON-
TROL connector and the SIGNAL I/O connector. If you do, the “REMOTE” mes-
sage will disappear and the START switch will be enabled.
• If the unconditional STOP function is enabled on the remote control box, the
tester will not enter the PROTECTION status when the plug of the box is con-
nected to the REMOTE CONTROL connector. Because the unconditional STOP
function is enabled on the HV test probe, the tester will not enter the PROTEC-
TION status when the probe is connected.
• Be sure to read section 6.3, “Remote Control” (Page 6-20) before remote-control-
ling the tester.
• To clear the PROTECTION status, press the STOP switch.
[5]
TEST VOLTAGE switch
The TEST VOLTAGE switch selects the test voltage range and switches the test
mode between AC and DC (TOS5050A has no DC test mode).
The TEST VOLTAGE switch is enabled when the tester is in the READY status
(“READY” is lit); it is disabled when the tester is in the TEST-ON status (“TEST”
is lit) or when the tester is delivering the result of PASS/FAIL judgment (“PASS” or
“FAIL” is lit).
Model
Selectable ranges and mode
TOS5051A
5 kVAC
2.5 kVAC
5 kVDC
2.5 kVDC
TOS5050A
5 kVAC
2.5 kVAC
-
-
WARNING
NOTE