PMX_MULTI
User’s Manual
73
Parallel Operation and Series Operation | Overview of Series Operation
Memory
.
The values set on CH2 (master) during series operation can be saved and recalled.
For the other channels of PMX32-3TR/PMX32-2QU, the values are shared between normal operation,
parallel operation, and series operation.
External control
.
To perform output on/off control, use pin 9 (CH2) and pin 13. Open pin 8 (CH1).
Alarms
Alarms that are detected on a channel are also detected during series operation
.
Clearing alarms
After resolving the cause of the alarm, press ALM CLR (SHIFT+SET), or turn the POWER switch off and
then on. If you do not resolve the cause of the alarm, an alarm will occur again.
If an alarm still occurs even after you have resolved all the causes of alarms, the product may be malfunc-
tioning. Stop using it immediately, and contact your Kikusui agent or distributor.