PLZ-U
Application Operation 5-19
■
Function for turning the load off in the OCP (CONFIG number 3)
Selects whether to turn off the load or limit the current when the OCP trips.
■
Function for turning the load off in the OPP (CONFIG number 4)
Selects whether to turn off the load or limit the power when the OPP trips.
■
Delayed load-on time setting (CONFIG number 5)
Set the delay time for turning on the load.
■
Last load (CONFIG number 6)
Set whether to keep the load turned on or turn the load off at the completion of the
sequence.
■
Polarity of the load-on external control (CONFIG number 7)
Set the polarity of the LOAD ON signal from the CH CONT connector.
Selectable Values
Description
OFF
Do not turn the load off (current limit operation)
ON
Turn the load off
Selectable Values
Description
OFF
Do not turn the load off (power limit operation)
ON
Turn the load off
Selectable Values
Description
0 ms to 1 s
Delay time
Selectable Values
Description
ON
Keep the load turned on at the completion of the sequence
OFF
Turn the load off at the completion of the sequence
Selectable Values
Description
L
Low active
H
High active