PLZ-U
Appendix A-23
A.6
Error Messages
A.6.1
Overview
Errors, when detected, are placed in the queue. The queue functions as a FIFO.
When the queue over
fl
ows, the last error in the queue is replaced by Error -350
(queue over
fl
ow). When all the errors are read from the queue, 0 (no error) is
returned for subsequent error queries.
When one of the following occurs, the error queue is cleared (section 11.4.3.4 in
IEEE 488.2).
•
A *CLS command is received.
•
The last item is read from the queue.
All negative values are reserved by the SCPI Standard. All errors speci
fi
c to the
PLZ-U are positive values.
A.6.2
A List of Errors
Command errors
An <error/event number> in the range [ -199, -100 ] indicates that an IEEE 488.2
syntax error has been detected by the instrument's parser. The occurrence of any
error in this class shall cause the command error bit (bit 5) in the event status regis-
ter to be set.
Error Code
Error Message Description
-100
Command error
Command error. This is the generic syntax error.
-101
Invalid character
A data element different from those that are allowed was encountered.
-102
Syntax error
An unrecognized syntax was encountered in the command string.
-103
Invalid separator
An invalid separator exists in the command string.
-104
Data type error
The parser recognized a data element different from the one allowed.
-105
GET not allowed
A Group Execute Trigger was received within a program message.
-108
Parameter not allowed
More parameters were received than expected for the header.
-109
Missing parameter
Fewer parameters were received than required for the header.
-110
Command header error
An error was detected in the header.
-120
Numeric data error
This error is generated when parsing a data element which appears to
be numeric, including the non-decimal numeric types.
-130
Suf
fi
x error
This error is generated when parsing a suf
fi
x.
-131
Invalid suf
fi
x
The suf
fi
x does not follow the syntax, or the suf
fi
x is inappropriate for
this device.
-134
Suf
fi
x too long
The suf
fi
x of the numeric element is too long.
-138
Suf
fi
x not allowed
A suf
fi
x was encountered after a numeric element which does not
allow suf
fi
xes.
Содержание PLZ-30F
Страница 32: ...2 18 Installation and Preparation PLZ U ...
Страница 58: ...4 14 Basic Operation PLZ U ...
Страница 144: ...6 52 Remote Control PLZ U ...
Страница 145: ...PLZ U Maintenance 7 1 Chapter 7 Maintenance This chapter explains the PLZ U maintenance 7 7 ...
Страница 186: ...A 26 Appendix PLZ U ...
Страница 190: ...I 4 Index PLZ U ...
Страница 191: ......