42
Operation Manual
PFX2731S
Specifications | Setting accuracy
Setting accuracy
Item
Specifications
Static
Constant current
charge/discharge
2 A range
Range
0.0000 A to 2.0000 A
Accuracy
1
1.
Ambient temperature range: 18 °C to 28 °C
±(0.15 % of s 1.0 mA)
Resolution
0.1 mA
20 A range
Range
0.000 A to 20.000 A
Accuracy
1
±(0.15 % of s 10.0 mA)
Resolution
1 mA
Constant voltage
charge/discharge
Range
-1.000 V to 6.000 V
Accuracy
1
±(0.05 % of s 1.2 mV)
Resolution
1 mV
Constant power
charge/
discharge
2
2.
The battery voltage is measured, and the control current (constant current control) is calculated from the set power value
through software calculation.
2 A range
Range
-0.100 W to 12.000 W
Accuracy
1,
3
3.
Battery voltage range: 1 V to 6 V or higher
±(0.5 % of s 0.01 W)
Resolution
1 mW
20 A range
Range
0.10 W to 120.00 W
Accuracy
1,
3
±(0.5 % of s 0.1 W)
Resolution
10 mW
Pattern
Constant current
pattern
2 A range
Range
-2.0000 A to 2.0000 A
(negative values are discharge currents)
Accuracy
1
±(0.15 % of s 1.0 mA)
Resolution
0.1 mA
20 A range
Range
-20.000 A to 20.000 A
Accuracy
1
±(0.15 % of s 10.0 mA)
Resolution
1 mA
Number of settings
100000 step (maximum number of steps)
Range
100 ms to 999.9 s (the time width per step)
Resolution
100 ms
Switching time
4
4.
Maximum time required for switching: charge -> discharge, discharge -> charge
100 ms max.
Constant power
pattern
2
2 A range
Range
-12.000 W to 12.000 W
(negative values are discharge power)
Accuracy
1,
3
±(0.5 % of s 0.01 W)
Resolution
1 mW
20 A range
Range
-120.00 W to 120.00 W
Accuracy
1,
3
±(0.5 % of s 0.1 W)
Resolution
10 mW
Number of settings
100000 step (maximum number of steps)
Range
100 ms to 999.9 s (the time width per step)
Resolution
100 ms
Switching time
4
100 ms max.