2
Features and Functions
64
Keysight U1731C/U1732C/U1733C User’s Guide
1
Press and hold the
key for more than 1 second to enter the open/short
CAL mode for the selected frequency and range.
2
Open/Short CAL prompts will be shown on the display. Follow the prompts for
open connector (
OPn
) or short connector (
SHor
) connection and press the
key.
The
CAL
annunciator on the upper right of the display will flash indicating that
the correction is in process.
3
After the open/short CAL is completed, the LCR meter will be restored to the
normal display and ready for normal usage.