From the LCR meter’s display, you can view the following statistical data for any
set of readings:
– Max: highest reading since the static recording mode was enabled
– Min: lowest reading since the static recording mode was enabled
– Avg: average or mean of all readings since the static recording mode was
enabled
– MaxMinAvg: present reading (actual input signal value)
Press the
key for more than 1 second to enter the static recording mode.
Figure 2-14
Using the Rec function
Press
again to cycle through the Max, Min, Avg, or MaxMinAvg (present) input
values.
To exit this mode, press and hold the
key for more than 1 second.
Rec
Hold
Rec
Hold
>1s
Rec
Hold
Rec
Hold
NOTE
– Static recording captures only stable values and updates the memory; it will
not record any overload ( ) value for any of the LCR functions. In addition,
the LCR meter will not record values below 50 counts in capacitance
measurement.
– Static recording is only available in manual ranging; therefore, activation
while in autoranging will automatically set the LCR meter to manual ranging.