Performance Tests and Calibration
7
Keysight U1401B User’s and Service Guide
147
Output performance verification
To verify the output functions of the Handheld Multi-Function Calibrator/Meter,
perform the verification tests listed in
for the recommended test equipment for verifying each function.
Table 7-4
Output performance verification tests
Step
Function
Recommended test equipment
and connection
Range or
parameter
U1401B
output
Nominal error
within 1 year
1
Turn the rotary switch to any
one of the
positions.
Connect the U1401B output
terminals to the 34405A
Multimeter.
±1.5000 V
–1.5 V
±0.75 mV
0 V
±0.3 mV
+1.5 V
±0.75 mV
±15.000 V
–15 V
±7.5 mV
0 V
±3 mV
+15 V
±7.5 mV
2
Turn the rotary switch to any
one of the
positions.
Connect the U1401B output
terminals to the 34405A
Multimeter.
±25.000 mA
–25 mA
±12.5 µA
+25 mA
±12.5 µA
3
Turn the rotary switch to any
one of the
positions.
Connect the U1401B output
terminals to the 53131A
Universal Counter.
Frequency
(10 kHz)
4.8 kHz
±0.25 Hz
Frequency
(1 kHz)
600 Hz
±0.04 Hz
Duty cycle
(0.39% to
99.60%)
5 V, 25%
@150 Hz
±0.203%
5 V, 75%
@ 150 Hz
±0.208%
Содержание U1401B
Страница 1: ...Keysight U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 10: ...10 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 61: ...Making Measurements 3 Keysight U1401B User s and Service Guide 61 Figure 3 1 DC voltage measurement ...
Страница 76: ...3 Making Measurements 76 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 107: ...Application Examples 5 Keysight U1401B User s and Service Guide 107 Figure 5 5 Zener diode test ...
Страница 128: ...5 Application Examples 128 Keysight U1401B User s and Service Guide Figure 5 17 Verifying a frequency transmitter ...