7
Performance Tests and Calibration
154
Keysight U1401B User’s and Service Guide
Output voltage calibration
Follow the steps below to perform calibration for the output voltage ranges and
values listed in
:
1
As you enter each calibration step, the primary and secondary displays show
the
output voltage value
and “-rdy-” respectively.
2
Press
OUTPUT
. The primary and secondary displays show the
output voltage
value
and “00000” respectively, which means the present output level is as
shown on the primary display.
3
Press
or
to adjust the output voltage until the multimeter reading is
the same as the value shown on the primary display.
4
Press
MODE
to enter the next calibration step.
At the end of the last calibration step, the primary display will show “PASS” after
the
MODE
button is pressed.
Table 7-6
Output voltage calibration steps
Voltage range
Calibration step
Output voltage value
1.5 V
1
+0.0000 V
2
+1.1000 V
3
–1.1000 V
15 V
4
+00.000 V
5
+11.000 V
6
–11.000 V
Содержание U1401B
Страница 1: ...Keysight U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 10: ...10 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 61: ...Making Measurements 3 Keysight U1401B User s and Service Guide 61 Figure 3 1 DC voltage measurement ...
Страница 76: ...3 Making Measurements 76 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 107: ...Application Examples 5 Keysight U1401B User s and Service Guide 107 Figure 5 5 Zener diode test ...
Страница 128: ...5 Application Examples 128 Keysight U1401B User s and Service Guide Figure 5 17 Verifying a frequency transmitter ...