Performance Tests and Calibration
7
Keysight U1401B User’s and Service Guide
145
2
(cont)
Turn the rotary switch to
.
Press
to select AC.
Connect the
Normal Hi-Low
output terminals of the
calibrator to the U1401B input
terminals.
500 mV
500 mVrms
@ 45 Hz
±3.7 mVrms
500 mVrms
@ 5 kHz
±3.7 mVrms
500 mVrms
@ 20 kHz
±7.7 mVrms
5 V
5 Vrms
@ 45 Hz
±37 mVrms
5 Vrms
@ 5 kHz
±37 mVrms
5 Vrms
@ 20 kHz
±77 mVrms
50 V
50 Vrms
@ 45 Hz
±0.37 Vrms
50 Vrms
@ 5 kHz
±0.37 Vrms
50 Vrms
@ 20 kHz
±0.77 Vrms
250 V
250 Vrms
@ 45 Hz
±1.95 Vrms
250 Vrms
@ 5 kHz
±1.95 Vrms
250 Vrms
@ 20 kHz
±3.95 Vrms
3
Turn the rotary switch to
.
Press
to select frequency.
Connect the
Normal Hi-Low
output terminals of the
calibrator to the U1401B input
terminals.
100 Hz
10 Hz
@ 16 mV
±5 mHz
100 kHz
20 kHz
@ 16mV
±7 Hz
200 kHz
200 kHz
@ 24 mV
±30 kHz
Table 7-3
Input performance verification tests (continued)
Step
Test Function
Connection to 5520A
Range
5520A output
Error from
nominal 1 year
Содержание U1401B
Страница 1: ...Keysight U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 10: ...10 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 61: ...Making Measurements 3 Keysight U1401B User s and Service Guide 61 Figure 3 1 DC voltage measurement ...
Страница 76: ...3 Making Measurements 76 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 107: ...Application Examples 5 Keysight U1401B User s and Service Guide 107 Figure 5 5 Zener diode test ...
Страница 128: ...5 Application Examples 128 Keysight U1401B User s and Service Guide Figure 5 17 Verifying a frequency transmitter ...