7
Performance Tests and Calibration
142
Keysight U1401B User’s and Service Guide
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures are listed below. If the exact instrument is not available,
substitute with another calibration standard of equivalent accuracy.
Table 7-1
Recommended test equipment
Application
Recommended equipment
Recommended accuracy requirements
DC Voltage
Fluke 5520A
<
1/5 instrument 1 year spec
DC Current
Fluke 5520A
<
1/5 instrument 1 year spec
Resistance
Fluke 5520A
<
1/5 instrument 1 year spec
AC Voltage
Fluke 5520A
<
1/5 instrument 1 year spec
AC Current
Fluke 5520A
<
1/5 instrument 1 year spec
Frequency
Keysight 33250A
<
1/5 instrument 1 year spec
Temperature
Fluke 5520A
<
1/5 instrument 1 year spec
Duty Cycle
Fluke 5520A
<
1/5 instrument 1 year spec
Diode
Fluke 5520A
<
1/5 instrument 1 year spec