Keysight P937xA PXIe Service Guide 95
General Product Information
Accessing the Keysight License Manager (KLM) and Installing Your License
Files
S
12
) with loads attached to both ports during the measurement calibration.
Isolation errors affect transmission measurements primarily where the
measured signal level is very low.
The measurements most affected by isolation error terms are DUTs with large
insertion loss. Since these terms are low in magnitude, they are usually noisy
(not very repeatable).
Typical Cause of Failure
A loose cable connection or leakage between components is the most likely
cause of isolation problems.
After verifying the cable and its connections, the analyzer requires repair. Refer
to
“Shipping Your Network Analyzer for Service or Repair” on page 22
.
Accessing the Keysight License Manager (KLM) and Installing Your
License Files
To start KLM:
1. On the PC controller: In Windows 10, click the
Search Windows
icon (
).
2. In the Search Windows box that opens, type: "Keysight License Manager"
and when the Keysight License Manager link appears under Best Match,
click on the link.
3. The dialog box illustrated in
is displayed.
Figure 6-3
Keysight License Manager Dialog Box
This section assumes you have completed the Getting Prepared section
(i.e., That you have already received your instrument’s license key files
from Keysight and downloaded them to a USB stick. And, that your USB
stick is inserted in your instrument’s PC controller.
Содержание P937 A Series
Страница 4: ...4 ...
Страница 8: ...8 Keysight P937xA PXIe Service Guide Contents ...
Страница 26: ...24 Keysight P937xA PXIe Service Guide General Product Information Keysight Support Services and Assistance 2 ...
Страница 62: ...60 Keysight P937xA PXIe Service Guide Replaceable Parts Listings Replaceable Parts Listings 4 ...
Страница 88: ...86 Keysight P937xA PXIe Service Guide Removal and Replacement Procedures Post Repair Procedures 5 ...
Страница 102: ...2 Keysight P937xA PXIe Service Guide ...