Keysight P937xA PXIe Service Guide 91
General Product Information
Error Terms
To access the error terms from the soft front panel, perform the following
steps:
1. On the P937xA software’s toolbar, click
Response > Cal > Cal Sets & Cal
Kits > Cal Set Viewer
.
The Cal Set Viewer tool bar appears directly above the trace window.
2. In the Cal Sets drop-down list, select the desired cal set or use the System
Default.
3. Click
Standards
to view the raw measurement data of the Standard (ONLY
available with Unguided Cal (not ECal or Guided Cal)), or click
Error Terms
to view the corrected error term data.
4. Use the Error Term box drop-down list to select an error term or standard to
view.
5. Select the Enable check box to view the data on the PC display.
6. Compare the displayed measurement trace to the equivalent data that
starts in
, or to previously measured data, or
to the uncorrected performance specifications listed in
7. Print numerical data or print a plot of the measurement results.