Keysight P937xA PXIe Service Guide 31
Tests, Adjustments, and Troubleshooting
ANSI/NCSL Z540.3-2006 and ISO/IEC 17025 Verification
ANSI/NCSL Z540.3-2006 and ISO/IEC 17025 Verification
To meet the criteria for ANSI/NCSL Z540.3-2006 and ISO/IEC 17025
verification, perform the preliminary checks and all performance tests without
stopping to repair or adjust.
1
Refer to
for test flow. Print data at the
completion of all the tests, even if you are aware that the network analyzer did
not pass. If there is a failure, complete the verification before you troubleshoot,
repair, and adjust. After the failure has been corrected, repeat the entire set of
performance tests and generate a new set of data.
Figure 3-1 ANSI/NCSL Z540.3-2006 and ISO/IEC 17025 Verification Flowchart
1. Stop only in case of a catastrophic failure or cable connector damage.