244 Keysight N9927-90001 User’s Guide
IQ Analyzer (IQA) – Option 351 (CPU2 Only)
I/Q Analyzer (IQA) Mode Settings
—
Res BW
(Spectrum only) – Sets the resolution bandwidth of the FFT
analysis. Changing this value changes the FFT window size, analysis
length, and the sweep time (data acquisition capture length). This
softkey is only available when IQA is set to Spectrum (Freq Domain).
When the resolution bandwidth value is set to Auto (default) the
FieldFox automatically sets the optimum resolution bandwidth value.
(Default: 200 kHz, Minimum: 0.1 Hz, maximum: 3 MHz).
—
Analysis Length
Sets the FFT analysis length that is the length of the
complex I/Q data points (I/Q–pairs) vector that is used for spectrum
analysis, expressed in the number of samples. (Couplings: In Auto,
this setting is directly coupled to the Res BW and to Span. Minimum:
8, Maximum: 524,288. Where each FFT length "point" is equal to 1 I
and 1 Q point. So, 237 is equal to 237 I and 237 Q points).
—
FFT Window
Sets the FFT filtering window (spectrum analysis
window) shape for the I/Q data. Default: Flat Top
Then select one of the following:
There is a significant difference between resolution BW in FFT
measurements and the resolution BW of a traditional "sweeping"
spectrum analyzer (SA) measurement. The resolution BW in FFT
measurements does not directly control the IF BW like a traditional SA
swept measurement. Resolution bandwidth is directly related to the
analysis length (window) and or data acquisition capture length (sweep
time).
FFT Window
Description
Uniform
No windowing
Triangular
Barlett windowing
Hanning
This is a Hann window, sometimes called Hanning or "raised
cosine".
Hamming
This window has lower side-lobe peaks than the Hann window,
but has worse main-lobe flatness and poor side-lobe roll-off.
Blackman
This window is similar to the three-term Blackman-Harris
window, but it has simpler fixed-form coefficients and better
side-lobe roll-off.
Blackman Harris 3
This is the three-term Blackman-Harris window. This has the
minimum side-lobe level for a three-term cosine window.
Blackman Harris
This is the four-term Blackman-Harris window. This has the
minimum side-lobe level for a four-term cosine window with
60 dB/decade roll-off.
Gaussian Top
This is a five-term cosine window that resembles a Gaussian
window.
Содержание N9913A
Страница 6: ...6 ...
Страница 56: ...56 Keysight N9927 90001 User s Guide CAT Cable and Antenna Test Mode 2 Port Insertion Loss Measurements ...
Страница 82: ...82 Keysight N9927 90001 User s Guide TDR Time Domain Reflectometry Option 215 TDR Measurement Settings ...
Страница 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 142: ...142 Keysight N9927 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Страница 262: ...262 Keysight N9927 90001 User s Guide IQ Analyzer IQA Option 351 CPU2 Only Other Features Available ...
Страница 286: ...286 Keysight N9927 90001 User s Guide Interference Analyzer SA Mode Option 236 Record Playback SA Option ...
Страница 292: ...292 Keysight N9927 90001 User s Guide Reflection Mode SA Models Option 320 Reflection Mode Settings ...
Страница 316: ...316 Keysight N9927 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 332: ...332 Keysight N9927 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 346: ...346 Keysight N9927 90001 User s Guide VVM Vector Voltmeter Mode Option 308 A B and B A Measurements ...
Страница 370: ...370 Keysight N9927 90001 User s Guide AM FM Metrics SA Only and Option 355 Interpreting AM FM Metrics Error Messages ...
Страница 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Страница 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Страница 528: ...528 Keysight N9927 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 544: ...544 Keysight N9927 90001 User s Guide Safety Considerations Certification and Compliance Statements ...
Страница 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Страница 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Страница 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Страница 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Страница 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Страница 555: ...555 Keysight Handheld Analyzers N99xxA User s Guide E Keysight Software End User Licensing Agreement EULA ...
Страница 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Страница 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Страница 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Страница 565: ......